Steris Autoclave Troubleshooting
The Steris Autoclave is a low-temperature sterile device used for surgical and diagnostic devices. Autoclaves are usually controlled by a microcomputer to ensure they are sterilized properly and can be managed and troubleshooted by a user. Troubleshooting procedures can be found with the provided manufacturer manual, as well as customer service lines for your device. There are also simple steps you can follow on your own in order to troubleshoot your autoclave.
Instructions
-
Troubleshoot an Autoclave
-
1
Clean the autoclave with distilled water. If you notice there is mineral buildup inside the autoclave, the safest product to use is distilled water with a clean towel; however, depending on the type of autoclave you own, the instruction or user manual may specify a cleaning agent to use.
-
2
Remove any items from the autoclave. It is possible there are too many items in the autoclave and that the steam is pushing out air too slowly for the temperature to reach its destination or remain constant.
-
-
3
Check for a leak in the valves. It is possible there is a leak in the valves of the autoclave. Call a service representative to find the valve and leak for your device.
-
4
Vent out the chamber and check to make sure there is no water left in the chamber.
-
5
Look at the counter and make sure there are not water droplets dripping from the door during the incubating cycle. If you notice water dripping, clean the door gasket. If the dripping continues, replace the door gasket.
Troubleshooting Power for Incubator
-
6
Turn the incubator to the "On" position. If the incubator does not turn on, make sure the power supply is plugged into the wall receptacle.
-
7
Reset the building circuit breaker and turn the incubator to the "On" position again.
-
8
Allow the incubator to run through a cycle. If the incubator turns on and shuts off during the cycle, check the connector cable on the incubator.
-
9
Contact the manufacturer if the power does not turn on or continues to shut off intermittently. The unit could be defective.
-
1
References
- Photo Credit power cable ends image by Georgios Alexandris from Fotolia.com