How to Troubleshoot an Agilent HPLC

How to Troubleshoot an Agilent HPLC thumbnail
Agilent HPLCs can analyze many substances.

A high-performance liquid chromotography (HPLC) machine analyzes laboratory samples by injecting the sample into the machine. As the continually flowing mobile phase (or moving phase) solution of the HPLC carries the sample over the HPLC column, the molecules in the sample become separated based on size and polarity. The individual molecules exit the column and cause an electrical charge of varying intensity, depending on the concentration of the molecule. A computer attached to the HPLC reads the electrical charge. This shows up on the monitor as a peak. The collection of peaks from the sample is called a chromatogram. To use an Agilent HPLC effectively, you must know how to troubleshoot problems.

Things You'll Need

  • Agilent HPLC
  • 50/50 aceotnitrile/water solution
  • New column (optional)
  • Chemical standard made according to your lab's protocol
  • HPLC syringe or automated sample loader
Show More

Instructions

  1. Pressure Problems

    • 1

      Check the pressure of the HPLC pump. The pressure will vary depending on your lab's HPLC setup, and you should know what the normal range is. If the pressure is higher or lower than normal, this indicates a problem.

    • 2

      Check the HPLC pump for leaks if the pressure is low. Leaks occur at any connector and might not be obviously dripping. Check for buildup of salts around any connector. This is a sign of a slow leak. Tighten any fittings or replace worn fittings to remedy low pressure problems.

    • 3

      Turn the pump down to zero pressure and remove the column if you experience high pressure problems. With the column off, turn the machine back up to the resting flow rate (the flow rate you leave the machine at overnight when not in use). If there is still pressure without the column, flush the pump with a 50/50 acetonitrile/water mix until the pressure goes down. If the pressure stays at zero without the column attached, recondition or replace the column per your lab protocol.

    Noise Problems

    • 4

      Turn on the computer for your HPLC and run the program that produces your chromatograms. If the baseline of the chromatogram is noisy (not smooth) without injecting any samples into the HPLC, it is either caused by dirty mobile phase or electrical interference. Make new mobile phase according to your lab protocol. Always use HPLC grade water for the mobile phase, and filter the mobile phase well. The detector of your HPLC must be keep closed at all times.

    • 5

      Inject your prepared standard into the HPLC. If there is noise only when a sample is injected, check that you are not injecting any air bubbles into the HPLC. Air bubbles result in large, jagged noise peaks. If you are certain you aren't injecting air bubbles, the problem is likely an unclean solution used to make the standard. Remake the solution in which you dilute your standard, then remake the standard.

    • 6

      Watch the baseline for noise that appears in a regular cycle. This is normally caused by an air bubble within the system. Turn the pump to zero pressure, remove the column, and use the purge function to force the bubble out. Reattach the column and turn the pump back to its running pressure. Avoid air bubbles by keeping the mobile phase at a constant temperature and degassing the mobile phase per your lab's protocol.

    • 7

      Check the baseline for continued noise. If the baseline shows a series of constant and jagged noise rather than a straight line, refresh your electrodes. Depending on the model of Agilent HPLC you have, the electrode configuration and maintenance can vary. Follow your lab's protocol for cleaning, polishing and refilling your HPLC's electrodes.

    Peak Problems

    • 8

      Inject the prepared standard into the HPLC. If the sample peak is not symmetrical, or if the peak splits (one peak appears as two), the column packing material has worn down or become separated. Refresh or replace your column per your lab's protocol.

    • 9

      Inject the prepared standard. If the chromatogram shows peaks that are smaller than normal, you have a sensitivity issue or your standards are old. Remake the standards and run them again. If the peaks are still small, increase your HPLC's sensitivity to obtain a larger peak. A decrease in sensitivity over time is normal and indicates you need to purchase new electrodes in the near future. Changing your mobile phase ingredients and regularly replacing the pump seals also help increase sensitivity.

    • 10

      Inject a prepared standard into the HPLC. If the peaks are broader than normal, this indicates that you need to recondition your column to repack the column packing material. Also, check that you are running the HPLC pump at the correct flow rate, since low pressure will cause peaks to spread out.

Tips & Warnings

  • Try one thing at a time to fix your HPLC. Each "fix" requires the machine to re-equilibrate before you can see if it helped. HPLC maintenance requires patience.

  • Always keep extra HPLC parts, such as seals and columns, on hand for HPLC maintenance.

  • Turn your pump to zero pressure any time you remove the column from the pump. A sudden decrease in pressure can destroy the pump seals.

  • Before you remove electrodes, turn the HPLCs power off to avoid shock.

Related Searches:

References

  • Photo Credit science image by peter Hires Images from Fotolia.com

Comments

You May Also Like

  • How to Troubleshoot a Double-Peak HPLC

    High performance liquid chromatography, or HPLC, is a technique used to separate compounds from a mixed solution. A metal tube filled with...

  • How to Troubleshoot a HPLC 1100

    HPLC stands for High Performance Liquid Chromatography. When troubleshooting an HPLC, first you need to isolate the problem. HPLC contains several parts...

  • How to Troubleshoot the HPLC Pump

    HPLC stands for high performance liquid chromatography. The basics of HPLC are that a solvent is forced through a chromatography column at...

  • How to Troubleshoot an HPLC

    High performance liquid chromatography, or HPLC, is a technique used to separate out compounds in a mixture through the use of a...

  • How to Select Columns in HPLC

    A high performance liquid chromatography, or HPLC, column consists of a metal cylinder packed with tiny silica beads used to separate compounds...

  • Advantages of Yams

    Advantages of Yams. There are hundreds of varieties of yams, and most of them come from Africa. Although the two are not...

  • What You Should Know About Ohio Land Contracts

    Ohio allows land contract real estate sales. The seller acts as the lender in a land contract sale, collecting the interest and...

  • Why Use Buffers in HPLC?

    High performance liquid chromatography (HPLC) is a common means of instrumental chemical analysis. A crucial component of any HPLC methodology is the...

  • How to Troubleshoot a HPLC System

    High Performance Liquid Chromatography is a method for separating a mixture of compounds in order to purify, identify, and quantify each component....

  • How to Reduce HPLC Solvent Peak

    High performance liquid chromatography (HPLC) analyzes samples you inject into an HPLC machine. A solvent called the mobile phase continually flows through...

  • How to Troubleshoot HPLC Columns

    The column of a high performance liquid chromatography (HPLC) machine is the stationary phase of the extraction process. The liquid--typically a buffer...

  • How to Measure Signal to Noise for HPLC

    High performance liquid chromatography, or HPLC, is a complex chemical method for separating the components of a liquid. HPLC systems first use...

  • How to Troubleshoot a Gilson HPLC

    Gilson provides high performance liquid chromatography (HPLC) systems worldwide for scientists in fields including biotech, food and beverage, pharmaceutical and ...

  • Power Amplifier Linearization Techniques

    Power Amplifier Linearization Techniques. There are many methods for power amplifier linearization, each method having its ups and downs. The best choice...

  • How to Validate HPLC Instruments

    High performance liquid chromatography, or HPLC, is a common method of instrumental chemical analysis. The method involves injecting a liquid sample into...

  • How to Calibrate HPLC Columns

    High-performance liquid chromatography (HPLC) has numerous applications in laboratory analysis. As a prelude to performing an actual analysis, it is important to...

  • How to Troubleshoot a Pixter Touch Screen

    There are a couple of options for you if you experience performance issues with your Pixter Color touch-screen device. Some issues you...

  • Normal Phase HPLC Columns

    High-performance liquid chromatography (HPLC) is a powerful method to rapidly separate liquids from a mixture. There are a number of different techniques,...

  • Strengths & Weaknesses of HPLC

    Strengths & Weaknesses of HPLC. HPLC, or high-performance liquid chromatography, is an analytical technique performed in many laboratories in the separation of...

  • How to Perform an HPLC

    Performing a HPLC (high performance liquid chromatography) test is a technical skill which should only be carried out by trained personnel. HPLC...

Related Ads

Featured